PCB-BA-e-0712
14
6.9 Underfloor weighing
Objects unsuitable for placing on the weighing scale due to size or shape may be
weighed with the help of the flush-mounted platform.
Proceed as follows:
• Turn off balance.
• Open the closing lid at the bottom of the balance.
• Carefully and completely hook in the hook for under floor weighing.
• Place balance over an opening.
• Suspend the item to be weighed from the hook and carry out weighing.
Fig. 1: Setup of balance for underfloor weighing
CAUTION
• Always ensure that all suspended objects are stable enough to hold the desired
goods to be weighed safely (danger of breaking).
• Never suspend loads that exceed the stated maximum load (max) (danger of
breaking)
Always ensure that there are no persons, animals or objects that might be damaged
underneath the load.
!
NOTICE
After completing the underfloor weighing the opening on the bottom of the balance
must always be closed (dust protection).