Front-Panel Menu Operation 3
Keysight 33210A Service Guide 43
To Read the Calibration Information
You can access the instrument’s calibration memory to read the calibration count
and calibration message.
Calibration Count You can query the instrument to determine how many
calibrations have been performed. Note that your instrument was calibrated
before it left the factory. When you receive your instrument, read the count to
determine its initial value. The count value increments by one for each calibration
point, and a complete calibration may increase the value by many counts.
Calibration Message The instrument allows you to store one message in
calibration memory. For example, you can store the date when the last calibration
was performed, the date when the next calibration is due, the instrument’s serial
number, or even the name and phone number of the person to contact for a new
calibration.
1 Select the Cal Info interface.
Press and then select the Cal Info softkey from the “Test/Cal” menu.
The first line in the display shows the calibration count.
The second line shows the calibration message.
The last line indicates the current version of the firmware.
The calibration information will time-out and disappear after a few seconds.
Select the Cal Info softkey to show the information again.
2 Exit the menu.
Press the DONE softkey.
NOTE
You can record a calibration message only from the remote interface and only
when the instrument is unsecured.
You can read the message from either the front-panel or over the remote
interface. You can read the calibration message whether the instrument is
secured or unsecured.