4 Calibration Procedures
68 Keysight 33210A Service Guide
Internal Timebase Verification
This test verifies the output frequency accuracy of the instrument. All output
frequencies are derived from a single generated frequency.
1 Connect a frequency counter as shown below (the frequency counter input
should be terminated at 50 Ω).
2 Set the instrument to the output described in the table below and measure the
output frequency. Be sure the instrument output is enabled.
3 Compare the measured frequency to the test limits shown in the table.
Keysight 33210A Measurement
Function Amplitude Frequency Nominal Error
Q Sine Wave 1.00 Vpp 10.000,000,0 MHz 10.000 MHz
± 200 Hz
[a]
[a] The error is ± 100 Hz within 90 days of calibration, or ± 200 Hz within one year.