4 Calibration Procedures
94 Keysight 33210A Service Guide
Low Frequency Flatness Adjustment
The Low Frequency Flatness adjustment calculates the flatness response of
3 attenuator paths with the Elliptical filter and 2 attenuator paths with the Linear
Phase filter.
1 Set the DMM to measure Vrms. Make the connections shown on page 92.
2 Use the DMM to measure the output voltage for each of the setups in the table
below.
3 Using the numeric keypad or knob, adjust the displayed voltage at each setup
to match the measured voltage. Select ENTER VALUE.
4 After performing setup 38:
a If your calibration procedures require you to verify the adjustment just
made, exit the calibration menu and perform “Low Frequency Flatness
Verification” on page 71.
Setup
Nominal Signal
Frequency Amplitude
29
[a]
[a] Constants are stored after completing this setup.
1 kHz 0.56 Vrms Flatness for 0dB, Elliptical Filter
30
[a]
100 kHz 0.56 Vrms Flatness for 0dB, Elliptical Filter
31
[a]
1 kHz 0.56 Vrms Flatness for 0dB, Linear Phase Filter
32
[a]
100 kHz 0.56 Vrms Flatness for 0dB, Linear Phase Filter
33
[a]
1 kHz 1.7 Vrms Flatness for +10dB, Elliptical Filter
34
[a]
100 kHz 1.7 Vrms Flatness for +10dB, Elliptical Filter
35
[a]
1 kHz 5.6 Vrms Flatness for +20dB, Elliptical Filter
36
[a]
100 kHz 5.6 Vrms Flatness for +20dB, Elliptical Filter
37
[a]
1 kHz 5.6 Vrms Flatness for +20dB, Linear Phase Filter
38
[a]
100 kHz 5.6 Vrms Flatness for +20dB, Linear Phase Filter