Calibration Procedures 4
Keysight 33210A Service Guide 85
Self-Test
Self-Test is performed as the first step to ensure the instrument is in working
order before beginning any additional adjustments.
1 Press on the front panel. Select Perform Cal on the “Test / Cal” menu.
Enter setup number “1” and select BEGIN.
2 If the instrument fails any self-test, you must repair the instrument before
continuing the adjustment procedures.
NOTE
Be sure to follow the requirements listed in “Test Considerations” on page 63
before beginning any adjustments.
Setup
1 Performs the Self-test. The Main Output is disabled during test.
NOTE
The self-test procedure takes approximately 15 seconds to complete.