Calibration Procedures 4
Keysight 33210A Service Guide 103
Calibration Errors
The following errors are failures that may occur during a calibration. System error
messages are described in the Keysight 33210A User’s Guide. Self-test error
messages are described on page 120 of this document
701: Calibration error; security defeated by hard ware jumper
The function generator’s calibration security feature has been disabled by
temporarily shorting the two “CAL ENABLE” pads on the internal circuit board as
described starting on page 79.
702: Calibration error; calibration memory is secured
A calibration cannot be performed when calibration memory is secured. See “To
Unsecure and Secure for Calibration” on page 44 for a procedure to enter the
security code from the front panel. Use the CAL:SEC:STAT ON command to enter
the security code using the remote interface.
703: Calibration error; secure code provided was invalid
Invalid security code specified with the CAL:SEC:STAT ON command.
706: Calibration error; value out of range
You have entered a value that was unexpected by the calibration firmware. For
example, if a number is expected such a 50.XX ohms, and you enter 10 ohms, that
number is outside the expected range of valid inputs.
707: Calibration error; signal input is out of range
Occurs during the ADC Adjustment, setup 6, if the 1 Volt input voltage is too high.
May also occur during self-calibration (setup 7), run self-test to diagnose cause of
problem.
707: Calibration error; cal edge time; rise time cal error
707: Calibration error; cal edge time; fall time cal error
707: Calibration error; cal edge time; default values loaded
Indicates a failure in the rise-time or fall-time circuitry has prevented calibration.
The edge-time was calibrated using default values, limiting accuracy. Service is
required to correct the problem and achieve design accuracy for the rise and fall
times.