4 Calibration Procedures
102 Keysight 33210A Service Guide
a If your calibration procedures require you to verify the adjustment just
made, exit the calibration menu and perform “+20 dB Range Flatness
Verification” on page 76.
b If you are making all the adjustments and then verifying the instrument’s
performance, verify the output specifications of the instrument using the
“Performance Verification Tests” on page 64.
You have now completed the recommended adjustment procedures. Verification
of the output specifications is recommended.