4 Calibration Procedures
60 Keysight 33210A Service Guide
Time Required for Calibration
The Keysight 33210A can be automatically calibrated under computer control. With
computer control you can perform the complete calibration procedure and
performance verification tests in approximately 30 minutes once the instrument is
warmed-up (see Test Considerations on page 63). Manual adjustments and
verifications, using the recommended test equipment, will take approximately 2
hours.
START
Incoming
Verification?
YES
NO
Perform
Adjustments
(approx 1 Hour)
Do Performance
Verification Tests
(approx 1 Hour)
Do Performance
Verification Tests
(approx 1 Hour)
DONE