EasyManua.ls Logo

Keysight Technologies 33210A - Db Range Flatness Adjustments

Keysight Technologies 33210A
141 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 Calibration Procedures
96 Keysight 33210A Service Guide
0 dB Range Flatness Adjustments
1 Connect the power meter as shown on page 98.
2 Use the power meter to measure the output amplitude for each of the setups
in the table below.
NOTE
Setup 39 establishes the power meter reference for all the remaining setups in
this table. You must always perform setup 39 before any of the following setups.
Setup
Nominal Signal
Frequency Amplitude
39
[a]
100 kHz 0.28 Vrms 2 dBm Power Meter Reference for 0dB Range
40
[a]
200 kHz 0.28 Vrms 2 dBm Flatness for 0 dB, Elliptical Filter
41
[a]
500 kHz 0.28 Vrms 2 dBm Flatness for 0 dB, Elliptical Filter
42
[a]
1.5 MHz 0.28 Vrms 2 dBm Flatness for 0 dB, Elliptical Filter
43
[a]
3 MHz 0.28 Vrms 2 dBm Flatness for 0 dB, Elliptical Filter
44
[a]
4 MHz 0.28 Vrms 2 dBm Flatness for 0 dB, Elliptical Filter
45
[a]
6 MHz 0.28 Vrms 2 dBm Flatness for 0 dB, Elliptical Filter
46
[a]
8 MHz 0.28 Vrms 2 dBm Flatness for 0 dB, Elliptical Filter
47
[a]
9.9 MHz 0.28 Vrms 2 dBm Flatness for 0 dB, Elliptical Filter
48
[a]
200 kHz 0.28 Vrms 2 dBm Flatness for 0 dB, Linear Phase Filter
49
[a]
500 kHz 0.28 Vrms 2 dBm Flatness for 0 dB, Linear Phase Filter
50
[a]
1.5 MHz 0.28 Vrms 2 dBm Flatness for 0 dB, Linear Phase Filter
51
[a]
3.0 MHz 0.28 Vrms 2 dBm Flatness for 0 dB, Linear Phase Filter
52 Setup not used for this instrument
53 Setup not used for this instrument
[a] Constants are stored after completing this setup.

Table of Contents

Other manuals for Keysight Technologies 33210A

Related product manuals