Calibration Procedures 4
Keysight 33210A Service Guide 75
6 Compare the measured output to the test limits shown in the table.
50 Ω Sine Wave +13.00 dBm 8.000 MHz 0 dB ± 0.3 dB
Q 50 Ω Sine Wave +13.00 dBm 10.000 MHz 0 dB ± 0.3 dB
Keysight 33210A Measurement
Output Setup Function Amplitude Frequency Nominal Error