Calibration Procedures 4
Keysight 34970A/34972A Service Guide 81
Input
Channel 209
Quick
Check
[a]
Error from Nominal
Function
[b]
Range 24 hour 90 day 1 year
Short
Short
Short
Short
Short
DC Volts
100 mV
1V
10V
100V
300V
Q
± 3.5 mV
± 6 mV
± 40 mV
± 600 mV
± 6 mV
± 4 mV
± 7 mV
± 50 mV
± 600 mV
± 9 mV
± 4 mV
± 7 mV
± 50 mV
± 600 mV
± 9 mV
Short
Short
Short
Short
Short
Short
Short
2-Wire
Ohms
[c]
and
4-Wire
Ohms
100 W
1 kW
10 kW
100 kW
1 MW
10 MW
100 MW
Q
± 3.5 mW
± 6 mW
± 50 mW
± 500 mW
± 10 W
± 100 W
± 10 kW
± 4 mW
± 10 mW
± 100 mW
± 1 W
± 10 W
± 100 W
± 10 kW
± 4 mW
± 10 mW
± 100 mW
± 1 W
± 10 W
± 100 W
± 10 kW
[a] Q: Quick performance verification test points.
[b] Select 6½ digit resolution.
[c] For 2-wire ohms, an additional 4W of error must be added.
NOTE
Zero offset calibration using a multifunction calibrator is NOT recommended.
The calibrator and cabling offset can be large and unstable causing poor offset
calibration of the internal DMM.

Table of Contents

Other manuals for Keysight Technologies 34972A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 34972A and is the answer not in the manual?

Keysight Technologies 34972A Specifications

General IconGeneral
BrandKeysight Technologies
Model34972A
CategorySwitch
LanguageEnglish

Related product manuals