Keysight E5260/E5270 Programming Guide, Edition 4 2- 43
Remote Mode Functions
Initial Settings
AV command parameter number=1, mode=0 AV
Current measurement range with pulse Compliance range RI
without pulse auto
Voltage measurement range with pulse Compliance range RV
without pulse auto
Sweep source parameters cleared WV, WSV, WI, WSI
Automatic abort function off WM
Output after sweep measurement Start value WM
Pulse source parameters cleared PV, PI
Pulse sweep source parameters cleared PWV, PWI
Pulse width 0.001 s PT
Pulse period 0.01 s PT
Search source parameters cleared BSV, BSSV, BSI,
BSSI, LSV, LSSV,
LSI, LSSV
Search monitor parameters cleared BGV, BGI, LGV,
LGI
Output after search measurement Start value BSM, LSM
Search measurement data Source output value only BSVM, LSVM
Quasi-pulse source parameters cleared BDV
Quasi-pulsed spot measurement mode Voltage BDM
Quasi-pulse settling detection interval Short BDM
Hold time 0 s WT, PT, BDT, BST,
LSTM
Delay time 0 s WT, PT, BDT, BST,
LSTM
Setup Item Initial Setting Commands