4- 22 Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
Table 4-8 Available Parameter Values
type mode N
0 0 Value that defines the number of averaging samples given
by the following formula. 1 to 1023. Default value is 1.
Number of averaging samples = N × initial averaging
where initial averaging is the number of averaging samples
automatically set by Keysight E5270B and you cannot
change.
1 Number of averaging samples. 1 to 1023. Default value is 1.
2 Value that defines the number of averaging samples given
by the following formula. 1 to 100. Default value is 1.
Number of averaging samples = N × 128
The Keysight E5270B gets 128 samples in a power line
cycle, repeats this for the times you specify, and performs
averaging to get the measurement data.
1 0 Value that defines the integration time given by the
following formula. 1 to 127. Default value is 6.
Integration time = N × initial integration time
where initial integration time is the integration time
automatically set by Keysight E5270B and you cannot
change.
1 Value that defines the integration time given by the
following formula. 1 to 127. Default value is 3.
Integration time = N × 80 μsec
2 Value that defines the integration time given by the
following formula. 1 to 100. Default value is 1.
Integration time = N / power line frequency