Keysight E5260/E5270 Programming Guide, Edition 4 4- 29
Command Reference
Command Reference
Remarks In the limit search mode, if search cannot find the search target and the following
two conditions are satisfied, the E5260/E5270 repeats the binary search between the
last source value and the source start value.
target is between the data at source start value and the last measurement data.
target is between the data at source stop value and the data at:
source value = | stop start | / 2.
If the search cannot find the search target and the following two conditions are
satisfied, the E5260/E5270 repeats the binary search between the last source value
and the source stop value.
target is between the data at source stop value and the last measurement data.
target is between the data at source start value and the data at:
source value = | stop start | / 2.