4- 58 Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
FL
This command sets the connection mode of a filter for each channel.
A filter is mounted on each module. It assures clean source output with no spikes or
overshooting.
Syntax FL mode[,chnum[,chnum...[,chnum]...]]
A maximum of eight channels can be set.
Parameters mode : Status of the filter. Integer expression.
0: Disconnect (initial setting).
1: Connect.
chnum : Channel number. The value must be slot number where the module has
been installed. Integer expression. 1 to 8 (1 and 5 are not available for
HPSMU). See
Table 4-1 on page 4-9.
If you do not specify chnum, the FL command sets the same mode for all channels.
Example
Statements
OUTPUT @E5270;"FL"
OUTPUT @E5270;"FL 0,1,3,5"