EasyManua.ls Logo

Keysight Technologies E5260 Series

Keysight Technologies E5260 Series
350 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4- 58 Keysight E5260/E5270 Programming Guide, Edition 4
Command Reference
Command Reference
FL
This command sets the connection mode of a filter for each channel.
A filter is mounted on each module. It assures clean source output with no spikes or
overshooting.
Syntax FL mode[,chnum[,chnum...[,chnum]...]]
A maximum of eight channels can be set.
Parameters mode : Status of the filter. Integer expression.
0: Disconnect (initial setting).
1: Connect.
chnum : Channel number. The value must be slot number where the module has
been installed. Integer expression. 1 to 8 (1 and 5 are not available for
HPSMU). See
Table 4-1 on page 4-9.
If you do not specify chnum, the FL command sets the same mode for all channels.
Example
Statements
OUTPUT @E5270;"FL"
OUTPUT @E5270;"FL 0,1,3,5"

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Related product manuals