Keysight E5260/E5270 Programming Guide, Edition 4 4- 71
Command Reference
Command Reference
58 : Returns the trigger settings:
[TGP port,terminal,polarity,type]
[;TGP port,terminal,polarity,type]
:
[;TGP port,terminal,polarity,type]
TGSI mode;TGXO mode;TGSO mode;TGMO mode<CR/LF^EOI>
59 : Returns the multi channel sweep source settings:
WNX n,chnum,mode,range,start,stop[,comp[,pcomp]]
[;WNX n,chnum,mode,range,start,stop[,comp[,pcomp]]]
:
[;WNX n,chnum,mode,range,start,stop[,comp[,pcomp]]]
<CR/LF^EOI>
If no multi channel sweep source is set, the query response is:
WNX<CR/LF^EOI>
60 : Returns the time stamp setting:
TSC enable<CR/LF^EOI>
61 : Returns the display settings:
RED enable;
KLC lock;
DFM format;
SPA1,param;
SPA2,param;
MPA param;
SCH chnum;
MCH chnum<CR/LF^EOI>
62 : Available only for the Keysight E5270B with HRSMU and ASU.
Returns the ASU connection path:
SAP chnum,path
[;SAP chnum,path]
:
[;SAP chnum,path]<CR/LF^EOI>
If no channel is connected to ASU.
SAP<CR/LF^EOI>