Keysight E5260/E5270 Programming Guide, Edition 4 4- 131
Command Reference
Command Reference
NOTE It is not easy to determine the best wait time. If you specify it too short, the
measurement may start before device characteristics stable. If too long, time will be
wasted.
The initial wait time may be too short for measurements of high capacitance or slow
response devices. Then set the wait time longer.
For measurements of low capacitance or fast response devices, if measurement
speed has top priority or is more important than reliability and accuracy, set the wait
time shorter.