5- 10 Keysight E5260/E5270 Programming Guide, Edition 4
Error Messages
Error Codes
303 Excess voltage in MPSMU.
Voltage that exceeds maximum voltage at the present current range was
detected by a MPSMU. All output switches were set to OFF.
305 Excess current in HPSMU.
Current that exceeds maximum current at the present voltage range was
detected by a HPSMU. All output switches were set to OFF.
307 Unsupported module.
This module is not supported by this firmware version. Until you
update the firmware, use the Keysight E5260/E5270 with this module
removed.
310 Interlock open operation error. Initialized.
Initialization was automatically performed because the E5260/E5270
failed to set its output to 0 V when the interlock circuit was opened in
the high voltage condition. Any module may be defective. Perform
self-test.
311 ASU control cable was connected/disconnected.
The E5270B must be turned off when the Atto Sense and Switch Unit
(ASU) is connected/disconnected.
603 Sweep and pulse channels must be different.
Set the sweep source and the pulse source to different channels for the
staircase sweep with pulsed bias measurement (MM5).
610 Quasi-pulse source channel must be set.
Before triggering the quasi-pulsed spot measurement, send the BDV
command to set the quasi-pulse source.
620 TGP specified incorrect I/O port.
Specify trigger input for the Ext Trig In port, or trigger output for the
Ext Trig Out port by the TGP command. Refer to
“TGP” on page
4-114.
621 Specify trigger input port for PAX/WSX.
No trigger input port was specified for the PAX or WSX command.
Specify the trigger input port, or set the port as the trigger input port.
Refer to
“TGP” on page 4-114 to set trigger port.