122 Keysight Agile Signal Generator Service Guide
Assembly Replacement
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Recommended torque for all screws is 9 in–lbs.
3. Recommended torque for all cable connectors is 9 in–lbs.
4. Replace the inner bottom rear cover and the instrument outer cover. Refer
to “Inner Bottom Covers” on page 101 and “Outer Cover” on page 99.
5. Perform the post-repair service procedure and performance tests that
pertain to this replacement procedure, see Chapter 4, “Post-Repair
Procedures and Operational Performance Verification.”.