EasyManuals Logo

Keysight Technologies N5193A User Manual

Keysight Technologies N5193A
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #53 background imageLoading...
Page #53 background image
Keysight Agile Signal Generator Service Guide 43
Troubleshooting
No Rear-Panel USB Activity
The rear-panel type Mini-B 5 pin USB connector is used for remotely
controlling the signal generator using SCPI commands.
1. Connect a USB cable between a controller PC that has been configured to
communicate through a USB cable and the signal generator. The PC
should detect the signal generator almost immediately.
2. If the signal generator is not detected:
a. Verify the rear-panel Mini-B USB cable is inserted deeply enough to
make proper contact.
b. If the rear-panel Mini-B USB cable is inserted deeply enough to
make proper contact, replace the A7 CPU assembly.

Table of Contents

Other manuals for Keysight Technologies N5193A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5193A and is the answer not in the manual?

Keysight Technologies N5193A Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5193A
CategoryPortable Generator
LanguageEnglish

Related product manuals