26 Keysight Agile Signal Generator Service Guide
Troubleshooting
A17 Agile Mod/Filter
AMF: Agile Mod Filter Self Tests
1001 - 1102 Replace A17. Go to replacement section page 119
A20 Output Leveling Module
HBD: High Band Detector Self Tests
1201 - 1203 Replace A20.
ā First go to page 158 in Chapter 4 to record the
Hardware ID for A20 adjustments.
ā Go to replacement section page 125
EAT: Electronic Attenuator Self Tests
1401 Replace A20.
ā First go to page 158 in Chapter 4 to record the
Hardware ID for A20 adjustments.
ā Go to replacement section page 125
A13 Rear Panel I/O Board
Rear Panel Board Self Tests
1301 - 1302 Replace A13. Go to replacement section page 137
Table 1-1 Self Test Failures
Self Test Failure Number Go to troubleshooting section: