Features and Functions 3
Keysight 34420A User’s Guide 101
– Front Panel operation (continued)
To set the delay to 0 seconds, select the “parameter” level of the TRIG DELAY 
command. Move the flashing cursor to the “units” location on the right side of 
the display. Press   until ZERO DELAY is reached, then press  .
– To select the automatic trigger delay, select the “parameter” level of the TRIG 
DELAY command. Move the flashing cursor to the “units” location on the right 
side of the display. Press   until AUTO DELAY is reached, then press  .
– Remote operation:
You can use the following command to set the trigger delay.
TRIGger:DELay {<seconds>|MIN|MAX}
You can use the following command to set an automatic trigger delay.
TRIGger:DELay:AUTO {OFF|ON}