3 Features and Functions
118 Keysight 34420A User’s Guide
Parity Selection (RS-232)
You can select the parity for RS-232 operation. The meter is configured for even
parity with 7 data bits when shipped from the factory.
Select one of the following: None (8 data bits), Even (7 data bits), or Odd (7
data bits). When you set the parity, you are indirectly setting the number of
data bits.
The parity selection is stored in non-volatile memory, and does not change
when power has been off or after a remote interface reset.
Front Panel operation: set the parity under the 4: PARITY command in the I/O
menu.
See also To Set the Parityon page 211.
Remote operation: use the following commands to set the parity:
SYStem:COMMunicate:SERial:PARity {EVEN|ODD|NONE}
If you change the parity over the remote interface, you may lose control of the
meter.