Application Programs 6
Keysight 34420A User’s Guide 237
HP BASIC Language Programs
All of the HP BASIC examples in this chapter were developed and tested on an HP
9000 Series 300 controller. Each device on the GPIB (IEEE-488) interface must
have a unique address. You can set the meter’s address to any value between 0
and 30. The current address is momentarily displayed on the front panel when you
turn on the meter.
The GPIB (IEEE-488) address is set to “22” when the meter is shipped from the
factory. The examples in this chapter assume a GPIB address of 22. When sending
a remote interface command, you append this address to the GPIB interface’s
select code (normally 7). Therefore, with an address of 22 and a select code of 7,
the combination is “722”.
HP BASIC/GPIB Program Example 1
This example demonstrates the general sequence of programming and the use of
the MEASure? command to set and obtain a 4-wire Ohms measurement.
10 !
20 ! This program configures the meter for a 4-wire ohms measurement on
30 ! Channel 1 using the MEAS command.
40 !
50 REAL Value,Rdg ! Dimension real variables
60 CLEAR 7 ! Clear interface - send "device clear"
70 ASSIGN @Meter TO 722 ! Assign I/O path to address 722
80 !
90 OUTPUT @Meter;"*RST" ! Reset the meter
100 OUTPUT @Meter;"*OPC?" ! Send "1" to output buffer when reset complete
110 ENTER @Meter;Value ! Enter value
120 !
130 ! Select 4-wire ohms on the 10 ohm range with the least resolution;
140 !
150 OUTPUT @Meter;"MEAS:FRES? 10,MAX"
160 ENTER @Meter;Rdg ! Enter reading
170 PRINT Rdg ! Print reading
180 END