4 Remote Interface Reference
156 Keysight 34420A User’s Guide
Special Resistance Measurement Commands
The following commands provide offset compensation and low power or voltage
limited resistance measurement capability.
[SENSe:]
FRESistance | RESistance
:OCOMpensated {OFF | ON}
:OCOMpensated?
This command enables or disables the offset compensated resistance
measurements (see page 82). Once enabled, offset compensation will apply to
both 4-wire and 2-wire resistance measurements.
[SENSe:]
FRESistance
:POWer:LIMit[:STATe] {OFF | ON}
:POWer:LIMit[:STATe]?
This command enables or disable the 4-wire low power resistance mode (see
page 82).
[SENSe:]
FRESistance
:VOLTage:LIMit[:STATe] {OFF | ON}
:VOLTage:LIMit[:STATe]?
:VOLTage:LIMit:VALue {<value> | MIN | MAX}
:VOLTage:LIMit:VALue? {MIN | MAX}
This command enables or disables the 4-wire voltage limited resistance mode
(see page 83). The voltage limit <value> parameter will be rounded to one of
three values:
20 mV,
100 mV, or
500 mV
MIN sets 20 mV and MAX sets 500 mV.