Features and Functions 3
Keysight 34420A User’s Guide 83
Voltage Limited Measurements
Voltage limited resistance measurements clamp the open circuit voltage to
predefined limits.
Voltage limits apply only to 4-wire ohms measurements.
Voltage limited resistance measurements are only available on the 10 W and
100 W ranges. If a manual range is selected in either 2-wire or 4-wire ohms,
and then Voltage Limited measurements are enabled, the range will be set to
the highest valid range (either 10 W or 100 W).
The open circuit voltage can be limited to 20 mV, 100 mV, or 500 mV.
Setting the Low Power measurement mode will cancel the Voltage Limited
mode.
Front panel operation: Enable or disable the voltage limit under the 5: LOW
VOLT W command in the MEASure menu.
Set the voltage limit under the 6: LoV LIMIT W command in the MEASure menu.
Remote operation: the following command enables or disable the low voltage
limit mode:
SENSe:FRESistance:VOLTage:LIMit {OFF|ON}
Set the limit with the following command:
SENSe:FRESistance:VOLTage:LIMit:VALue {<value>|MIN|MAX}