Remote Interface Reference 4
Keysight 34420A User’s Guide 203
Using Device Clear to Halt Measurements
Device clear is an IEEE-488 low-level bus message which can be used to halt
measurements in progress. Different programming languages and IEEE-488
interface cards provide access to this capability through their own unique
commands. The status registers, the error queue, and all configuration states are
left unchanged when a device clear message is received. Device clear performs
the following actions.
All measurements in progress are aborted.
The meter returns to the trigger “idle state”.
The meter’s input and output buffers are cleared.
The meter is prepared to accept a new command string.
For RS-232 operation, sending the control-C character will perform the
equivalent operations of the IEEE-488 device clear message. The meter’s DTR
(data terminal ready) handshake line will be true following a device clear
message. See “DTR / DSR Handshake Protocol” on page 194 for further
details.