Remote Interface Reference 4
Keysight 34420A User’s Guide 167
Triggering
See also Triggering” on page 93 in Chapter 3.
First-time SCPI users, see page 198.
The meter’s triggering system allows you to generate triggers either manually or
automatically, take multiple readings per trigger, and insert a delay before each
reading. Normally, the meter will take one reading each time it receives a trigger,
but you can specify multiple readings (up to 50,000) per trigger. You can also set
an infinite trigger.
Triggering the meter from the remote interface is a multi-step process that offers
triggering flexibility.
First, you must configure the meter for the measurement by selecting the
function, range, resolution, etc.
Then, you must specify the source from which the meter will accept the
trigger. The meter will accept a software (bus) trigger from the remote
interface, a hardware trigger from the rear-panel Ext Trig (external trigger)
terminal, or an immediate internal trigger.
Then, you must make sure that the meter is ready to accept a trigger from the
specified trigger source (this is called the wait-for-trigger state).
The diagram on the next page shows the meter’s triggering system.