4 Remote Interface Reference
184 Keysight 34420A User’s Guide
The questionable data event register is cleared when:
You execute a *CLS (clear status) command.
You query the event register using STATus:QUEStionable:EVENt?.
The questionable data enable register is cleared when
You turn on the power (*PSC does not apply).
You execute the STATus:PRESet command.
You execute the STATus:QUEStionable:ENABle 0 command.
The Operational Status and Condition Registers
The operation status register provides information about the operation of the
meter. Bits in the operational status event register are latched from changes in the
bits in the operational status condition register.
The meter only uses a single bit (bit 8) in these registers. This bit indicates the
settling state of the digital filter. If the digital filter is enabled, bit 8 is set to 1 to
indicate that the filter has settled. See page 70 for information about the
operation of the digital filter.
You can read the operational status condition register to determine the digital
filter state on a per measurement basis. You can read the operational status event
7 Not Used 128 Always set to 0.
8 Not Used 256 Always set to 0.
9 Ohms Overload 512 Range overload on 2-wire or 4-wire Ohms
10 Not Used 1024 Always set to 0.
11 Not Used 2048 Always set to 0.
12 Not Used 4096 Always set to 0.
13 Not Used 8192 Always set to 0.
14 Not Used 16384 Always set to 0.
15 Not Used 32768 Always set to 0.
Bit Decimal Value Definition