Features and Functions 3
Keysight 34420A User’s Guide 111
Chart Output (Analog Output)
You can use the Chart Output (Analog Output) connector on the rear panel to run
a strip chart recorder or similar instrument. The connector provides an output
voltage proportional to the measured voltage.
The output can range from –3.00 V to +3.00 V.
When chart output is disabled (Off), the output is set to 0 V.
The output impedance is 1 kW.
The output voltage is related to the input signal as follows:
Note that reading includes the results of measurement unit conversion,
filtering, null, and any math operation performed on the measurement.
The chart output state (including the span and null values) is stored in
non-volatile memory.
span cannot have a negative value.
The chart output has a voltage “rollover” to prevent the loss of data, see
page 113.
Examples:
See page 48 for a menu procedure to enter a number.
1 To set the chart span so that a ±1 mV input signal produces a ±1.0 Volt chart
output:
When chart null is set to 0, the value can be considered the gain and
chart output = gain * reading.
reading chart null()
span
----------------------------------------- chart output voltage=
0.001 0.0()
span
-------------------------1.00=
span
0.001
1.00
-----------0.001==