3 Features and Functions
84 Keysight 34420A User’s Guide
Temperature Measurement Configuration
This section contains information to help you configure the meter for making
temperature measurements. To measure temperature, you will need a
temperature transducer. Transducer descriptions and specific comments about
their usage are given on page 288. The meter supports three general categories of
transducer:
RTD (Resistive Temperature Device)
THERM (Thermistor)
T/C (Thermocouple)
The meter supports specific types of transducers in each category:
Once you set the measurement units, probe type, and transducer type, the setting
is stored in non-volatile memory. This makes the last probe type used the default
probe type.
Measurement Units
The meter can report temperature in °C (Celsius), °F (Fahrenheit), or K
(Kelvins).
The measurement units set are stored in non-volatile memory.
Front Panel Operation: Set the units under the 2: UNITS command in the
TEMPerature menu.
Remote operation: Set the units using the following command:
UNIT:TEMPerature {C|F|K}
RTD’s
R
0
= 4.9 W to 2.1 kW
Thermistor Thermocouples
a = 0.00385 (DIN/IEC 751)
a = 0.00391
5k W B, E, J, K, N, R, S, T