Remote Interface Reference 4
Keysight 34420A User’s Guide 163
INPut:FILTer
:STATe {OFF | ON}
:STATe?
These commands enable or disable the filter state and query the filter state.
INPut:FILTer
:TYPE {ANALog | DIGital | BOTH}
:TYPE?
These commands set the filter type and query the filter type.
BOTH enables the digital and analog filters.
INPut:FILTer
:DIGital:RESPonse {SLOW | MEDium | FAST}
:DIGital:RESPonse?
These commands set the averaging for the digital filter and query the digital filter
setting. SLOW sets a 100 reading average, MEDium sets a 50 reading average,
and FAST sets a 10 reading average.
INPut:FILTer
:DIGital:PRECharge {OFF | ON}
:DIGital:PRECharge?
These commands enable or disable the digital filter precharge and query the
digital filter precharge setting. Disabling the digital filter precharge prevents the
filter from resetting based upon the measured value (see page 70 in Chapter 3).