5 Error Messages
226 Keysight 34420A User’s Guide
-214 Trigger deadlock
A trigger deadlock occurs when the trigger source is BUS and a
READ? command is received.
-221 Settings conflict
This error can be generated in one of the following situations:
You sent a
CONFigure or MEASure command with autorange
enabled and with a fixed resolution. Example:
CONF:VOLT:DC
DEF,0.1
You turned math on (CALC:STAT ON) and then changed to a math
operation that was not valid with the present measurement
function.
-222 Data out of range
A numeric parameter value is outside the valid range for the
command. Example:
TRIG:COUN -3
-223 Too much data
A character string was received but could not be executed
because the string length was more than 11 characters. This
error can be generated by the
CALibration:STRing and
DISPlay:TEXT commands.
-224 Illegal parameter value
A discrete parameter was received which was not a valid choice
for the command. You may have used an invalid parameter
choice. Example:
CALC:FUNC SCALE (SCALE is not a valid choice).
-230 Data corrupt or stale
A
FETCh? command was received but internal reading memory
was empty. The reading retrieved may be invalid.
-330 Self-test failed
The meter’s complete self-test failed from the remote interface
(
*TST? command). In addition to this error, more specific
self-test errors are also reported. See also “Self-Test Errors” on
page 230.