noise measurements for the 1 mV range with the low thermal shorting plug 
applied to the input terminals. Other functions and ranges will exhibit similar 
noise reduction characteristics versus NPLC.
For example, if you set the NPLC to 20 and enable the digital filter set to FAST (10 
reading average), the equivalent NPLC is 200. At an equivalent NPLC of 200, the 
reading rate is approximately 6 readings per minute (digital filter fully settled) and 
the typical noise level expected is 1.6 nVrms.
Alternately, if you take readings every 5 minutes, you can set an equivalent NPLC 
of 5000 to minimize the noise in the measurement. An equivalent NPLC of 5000 
could be obtained by setting the integration time to 100 NPLC and enabling the 
digital filter set to MEDIUM (50 reading average). At this equivalent NPLC, a 
typical noise level of 0.35 nVrms can be expected.
– Front Panel operation: Integration time can be set indirectly when you select 
the number of digits (see page 54).
You can also set the integration time in the MEASurement menu with the 
2: INTEGRATE command.
Enable the digital filter and set the filter reading average as described on 
page 68.