Keysight 3458A Quick Reference Guide 15
Keysight 3458A Quick Reference Guide 15
ACDCI, ACDCV, ACI, ACV
Refer to the FUNC command.
ADDRESS value
Sets the multimeter's GPIB address (from the
front panel only). The address is stored in
continuous memory and is not lost when power is
removed.
value: The value parameter is an integer from 0 to
31.
Power-on value = previously stored address
(factory setting = 22).
Default value = none: parameter required.
APER [aperture]
Aperture. Specifies the A/D converter integration
time in seconds.
aperture: Specifies the A/D converter's
integration time and overrides any previously
specified integration time or resolution. The valid
range for aperture is 0 - 1s in increments of 100
ns. (Specifying a value <500 ns selects minimum
aperture which is 500 ns.)
Power-on aperture = is determined by the
power-on value for NPLC which specifies an
integration time of 166.667 ms for a 60 Hz power
line frequency, or 200 ms for a power line
frequency of 50 Hz or 40O Hz.
Default aperture = 5OO ns.