EasyManua.ls Logo

Keysight 3458A - Before Applying Power Procedures

Keysight 3458A
61 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Summary
Keysight 3458A Quick Reference Guide 29
Keysight 3458A Quick Reference Guide 29
Default control = ON.
ISCALE?
Integer Scale Query. Returns the scale factor for
readings output in the SINT or DINT formats.
LEVEL [percentage][,coupling]
The LEVEL command specifies the level triggering
voltage (as a percentage of the present range)
and the coupling (AC or DC) for level triggering. A
level trigger event occurs when the input signal
reaches the specified voltage on its
positive-going or negative-going slope as
specified by the SLOPE command.
percentage: Specifies the percentage of the
present range for level triggering. The valid range
for this parameter is -500% to +500% in 5% steps
for direct or sub-sampling or -120% to 120% in
1% steps for DC voltage.
Power-on percentage = 0% (0V).
Default percentage = 0% (0V).
coupling = DC (1), AC (2)
Power-on coupling = AC.
Default coupling = AC.
LFILTER [control]
Level Filter. Enables or disables the level filter
function. When enabled. the level filter function
connects a single pole low-pass filter circuit to
the input of the level-detection circuitry. The
low-pass filter has a 3-dB point of 75 kHz and

Table of Contents