EasyManua.ls Logo

Keysight 3458A - MSIZE [reading_mem][,subprogram_mem]; NDIG [value]

Keysight 3458A
61 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Summary
Keysight 3458A Quick Reference Guide 33
Keysight 3458A Quick Reference Guide 33
PERC (10), PFAIL (11), RMS (12), SCALE (13),
STAT (14), CTHRM2K (16), CTHRM10K (17),
FTHRM2K (18), FTHRM10K (19), CRTD85 (20),
CRTD92 (21), FRTD85 (22), FRTD92 (23)
Power-on operation_a, operation_b = OFF, OFF.
Default operation_a, operation_b = OFF, OFF.
Power-on register values = all registers are set to
0 with the following exceptions:
MSIZE [reading_mem][,subprogram_mem]
Memory size. On a previous Agilent multimeter,
the MSIZE command was used to clear all
memory and allocate memory space for readings,
subprograms, and state storage. The 3458A
accepts the MSIZE command to maintain
language compatibility, but performs no action
since the 3458A's memory allocations are
predefined and cannot be changed. The MSIZE?
query command, however, is used to determine
the total reading memory and
the largest unused block of subprogram/state
memory.
NDIG [value]
Number of Digits. Designates the number of
display digits.
DEGREE = 20 REF = 1
SCALE = 1 RES = 50
PERC = 1

Table of Contents

Other manuals for Keysight 3458A

Related product manuals