GPIB Command Summary
58 Keysight 3458A Quick Reference Guide
58 Keysight 3458A Quick Reference Guide
REMOTE
Sets the GPIB REN line true.
SPOLL (Serial Poll)
The SPOLL command, like the STB? command
(multimeter command set), returns a number
representing the set bits in the status register
(status byte). The returned number is the
weighted sum of all set bits.
The status register conditions and their weights
are:
TRIGGER (GET)
If triggering is armed (see TARM command), the
TRIGGER command (Group Execute Trigger)
triggers the multimeter once, and then holds
triggering.
Decimal
Weight
Bit
Number
Enables Condition
1 0 Program Memory Execution Complete
2 1 Hi or Lo Limit Exceeded
4 2 SRQ Command Executed
83Power-On SRQ
16 4 Ready for instructions
32 5 Error (See ERRSTR?)
64 6 Service Requested
128 7 Data AvaiLable