Command Summary
Keysight 3458A Quick Reference Guide 55
Keysight 3458A Quick Reference Guide 55
control = OFF (0), ON (1)
Power-on control = OFF.
Default control = OFF.
TEMP?
Temperature Query. returns the multimeter's
internal temperature in degrees Centigrade.
TERM [source]
On previous Keysight multimeters, the TERM
command internally connected or disconnected
the multimeter's input terminals. The 3458A
accepts the TERM command to maintain
language compatibility with these multimeters,
but does not respond since the 3458A's input
terminals are controlled with a manually operated
switch.
Power-on source = none.
Default source = FRONT.
TEST
Causes the multimeter to perform a series of
internal self-tests.
TIMER [time]
The TIMER command defines the time interval for
the TIMER sample event in the NRDGS command.
When using the TIMER event. the time interval is
inserted between successive measurements.