Command Summary
18 Keysight 3458A Quick Reference Guide
18 Keysight 3458A Quick Reference Guide
CALL [name]
Call Subprogram. Executes a previously stored
subprogram.
name: A subprogram name may contain up to 10
characters. The name can be alpha,
alphanumeric, or an integer in the range of 0 to
127.
Power-on name = none.
Default name = 0.
CALNUM?
Calibration Number Query. Returns an integer
indicating the number of times the multimeter
has been calibrated.
CALSTR string[,security_code]
Calibration String (remote only). Stores a string
in the multimeter's nonvolatile calibration RAM.
string: This is the alpha/numeric message that
will be appended to the calibration RAM. The
string
parameter must be enclosed in single or double
quotes. Maximum string length = 75 characters
(the quotes enclosing the string are not counted
as characters).
security_code: When the calibration RAM is
secured (SECURE command) you must include
the security_code in order to write a message to
the calibration RAM.