GPIB Command Summary
Keysight 3458A Quick Reference Guide 57
Keysight 3458A Quick Reference Guide 57
GPIB Command Summary
ABORT 7 (IFC)
Clears the multimeter's interface circuitry.
CLEAR (DCL or SDC)
Clears the multimeter, preparing it to receive a 
command. The CLEAR command does the 
following:
Clears the output buffer.
Clears the input buffer.
Aborts subprogram execution.
Clears the status register (bits 4, 5, and 6 are not 
cleared if the condition(s) that set the bit(s) still 
exist).
Clears the display.
Disables triggering (the previous triggering mode 
can be resumed by sending any multimeter 
command).
LOCAL(GTL)
Removes the multimeter from the remote state 
and enables its keyboard (provided the keyboard 
has not been disabled with the multimeter’s 
LOCK command).
LOCAL LOCKOUT (LLO)
Disables the multimeter's LOCAL key.