Command Summary
Keysight 3458A Quick Reference Guide 37
Keysight 3458A Quick Reference Guide 37
%_resolution: Specifies the digits of resolution
and the gate time as shown below:
Power-on %_resolution = not applicable.
Default %_resolution = .00001
PRESET [type]
Configures the multimeter to one of three
predefined states.
type: Specifies the NORM, FAST, or DIG preset
state (the numeric query equivalents of these
parameters are 1, 0, and 2, respectively).
Power-on type = not applicable.
Default type = NORM.
PRESET NORM is similar to RESET but optimizes
the multimeter for remote operation. Executing
PRESET NORM executes the following
commands:
%_resolution
Parameter
Selects
Gate Time
Digits of
Resolution
.00001 1 s 7
.0001 100 ms 7
.001 10 ms 6
.01 1 ms 5
. 1 100 µs 4