record: Designates the record from which to recall 
readings. Records correspond to the number of 
readings specified by the NRDGS command. For 
example, if NRDGS specifies three readings per 
trigger, each record will contain three readings.
Power-on record = none.
Default record = 1
RQS [value]
Request Service. Enables one or more status 
register conditions. When a condition is enabled 
and that condition occurs, it sets the GPIB SRQ 
line true.
value: You enable a condition by specifying its 
decimal weight as the value parameter. For more 
than one condition, specify the sum of the 
weights. Refer to the STB? command for a table 
showing the conditions and their weights.
Power-on value: If Power-On SRQ was enabled 
when power was removed, value = 8: otherwise, 
value = 0
Default value = 0 (no conditions enabled).
RSTATE [name]
Recall State. Recalls a stored state from memory 
and configures the multimeter to that state. 
States are stored using the SSTATE command.
name: A state name may contain up to 10 
characters. The name can be alpha, 
alphanumeric, or an integer in the range of 0 to 
127.
Power-on name = none.