Command Summary
Keysight 3458A Quick Reference Guide 23
Keysight 3458A Quick Reference Guide 23
EMASK [value]
Error Mask. Enables certain error condition(s) to 
set the error bit (bit 5) in the status register.
value: To enable an error condition, specify its 
decimal weight as the value parameter. To enable 
more than one error condition, specify the sum of 
the weights. The error conditions and their 
weights are:
Power-on value = 32767 (all enabled).
Weighted
Value
Bit
Number
Error Conditions
1 0 Hardware, see AUXERR?
21Calibration
4 2 Trigger too fast
83Syntax error
16 4 Not allowed from remote
32 5 Undefined parameter
64 6 Parameter out of range
128 7 Memory Error
256 8 Destructive overload
512 9 Out of calibration
1024 10 Calibration required
2048 11 Settings conflict
4096 12 Math error
8192 13 Subprogram error
16384 14 System error