Command Summary
54 Keysight 3458A Quick Reference Guide
54 Keysight 3458A Quick Reference Guide
#_samples: Specifies the number of samples to
be taken. The valid range for this parameter is 1
to 1.67E+7.
Power-on #_samples = 1024
Default #_samples = 1024
T [event]
T is an abbreviation for the TRIG command. Refer
to the TRIG command for more information.
TARM [event][,numher_arms]
Trigger Arm. Defines the event that
enables (arms) the trigger event (TRIG
command). You can also use this command to
perform multiple measurement cycles.
event = AUTO (1). EXT (2). SGL (3). HOLD (4). SYN
(5)
Power-on event = AUTO.
Default event = AUTO.
number-arms: Specifies multiple trigger arming
(the TARM event must be SGL). The valid range is
0-2.1E+9.
Power-on number_arms = 1 (multiple arming
disabled)
Default number_arms = 1 (multiple arming
disabled)
TBUFF [control]
Trigger Buffer. Enables or disables the
multimeter's external trigger buffer.