Command Summary
50 Keysight 3458A Quick Reference Guide
50 Keysight 3458A Quick Reference Guide
SSAC [max._input] [,%_resolution]
or
SSDC [max._input] [,%_resolution]
Sub-Sampling. Configures the multimeter for
sub-sampled voltage measurements (digitizing).
The SSAC function measures only the AC
component of the input waveform. The SSDC
function measures the combined AC and DC
components of the waveform. Otherwise, the two
functions are identical. The input signal must be
periodic (repetitive) for sub-sampled
measurements. Sub-sampled measurements use
the track/hold circuit (2 nanoseconds aperture)
and a wide bandwidth input path (12 MHz
bandwidth).
max._input: Selects the measurement range. (See
the RANGE command for details).
Power-on max._input = not applicable.
Default max._input = 10V.
%_resolution: Is ignored when used with the
SSAC or SSDC command.
SSPARM?
Sub-Sampling Parameters Query. Returns the
parameters necessary to reconstruct a
sub-sampled waveform (SSAC or SSDC
command) when the samples are sent directly to
the GPIB output buffer. (Reconstruction is
automatic when the samples are sent directly to
reading memory).