Command Summary
22 Keysight 3458A Quick Reference Guide
22 Keysight 3458A Quick Reference Guide
DISP [control] [,message]
Display. Enables or disables the multimeter’s
display, and may also be used to send a message
to the display or to clear the display.
control = OFF (0), ON (1), MSG (2), CLR (3)
Power-on control = ON.
Default control = ON.
message = the message to be displayed.
DSAC [max._input] [,%_resolution]
or
DSDC [max._input] [,%_resolution]
Direct-Sampling. Configures the multimeter for
direct-sampled measurements (digitizing). The
DSAC function measures only the AC component
of the input waveform. The DSDC function
measures the combined AC and DC components.
Otherwise. the two functions are identical. The
DSAC and DSDC functions use the track/hold
circuit (2 nanosecond aperture) and a wide
bandwidth input path (12 MHz bandwidth).
max._input: Selects the measurement range. (See
The RANGE command for details.)
Power-on max._input = not applicable.
Default max._input = 10V.
%_resolution: Is ignored when Used with the
DSAC or DSDC command.