Command Summary
Keysight 3458A Quick Reference Guide 43
Keysight 3458A Quick Reference Guide 43
For ACI or ACDCI:
For DSAC or DSDC:
max._input
Parameter
Selects Range
Full
Scale
–1 or AUTO Autorange
0 to .120E-6 100 µA 120 µA
>120E-6 to 1.2E-3 1 mA 1.2 mA
>1.2E-3 to 12E-3 10 mA 12 mA
>12E-3 to 120E-3 100 mA 120 mA
>120E-3 to 1.2 1 A 1.05 A
max._input
Parameter
Selects Range
Full Scale
SINT
format
DINT
format
0 to .012 10 mV 12 mV 50 mV
>.012 to .120 100 mV 120 mV 500 mV
>.120 to 1.2 1 V 1.2 V 5.0 V
>1.2 to 12 10 V 12 V 50 V
>12 to 120 100 V 120 V 500 V
>120 to 1E3 1000 V 1050 V 1050 V