230 Chapter 13
Assembly Replacement Procedures
Instrument Outer Case
Instrument Outer Case
CAUTION If the instrument is placed on its face during any of the following procedures, be
sure to use a soft surface or soft cloth to avoid damage to the front panel, keys, or
input connector.
NOTE If the analyzer has Option PRC, Portable Configuration, refer to the “Option PRC”
section on page 232 to remove the outer case.
Standard Instrument
Removal
1. Disconnect the instrument from ac power.
2. Refer to Figure 13-3. Using the T-20 driver, remove the four screws (two on
each side) (1) that attach the handle strap (2) on each side of the instrument.
3. Using the T-20 driver, remove the four screws (including washers) (3) that
hold the rear feet (4) in place.
4. Pull the instrument cover (5) off towards the rear of the instrument.