Assembly Replacement Procedures
Input Connector Assembly
Input Connector Assembly
Removal
1. Remove the instrument outer case. Refer to the Instrument Outer Case removal
procedure.
2. Remove the Front Frame Assembly. Refer to the Front Frame Assembly
removal procedure.
3. Refer to Figure 13-28. Disconnect the semi-rigid cable W8 from the Input
Connector Assembly.
Figure 13-28 Input Connector Assembly Removal
4. Remove the two screws (1) that attach the Input Connector Assembly to the
chassis. The connector assembly can now be lifted from the chassis.
2
W8
1